FAT Issues
TheFactoryAcceptanceTestcanbethefirsttimethemeterisrunafterbeingdespatchedfromthe
factory.Herearesomecommonproble msencounteredandtheirresolution:
1 ThemostfrequentlyusedFATfluidiswater.Thiswon’tdamagethemetersatall,butit
doeshaveaverylow
viscosity.Therearetwosignificanteffects.Firstly,themeterwill
probablynotmeasureaslowastheflowratesachievedattimeoffactorycalibration.
Thecalibrationviscosityistakentobetheviscosityofnormaloperation.Secondly,ifit
doesoperateatthelowestflows,thentheaccuracymay
begreatlyreduced.Itislikely
tounderreadbyupto30%.ThesolutionistochangetheFATfluid.Theadditionof
glycolorreplacementbyglycolfortheFATisquitecommon.
2 Dirtinthesystemishardtoavoid.Evenwithfiltrationat40or100
micronsitisstill
quiteeasyforpotentiallyblockingparticlestobeheldinthepiping,evenafterflushing.
Wewouldrecommendthemeteristakenoutoflinebeforepipeflushing.Itemslike
PTFEtapeandtheremnantsofthreadlockingcompoundswillpreventtherotorfrom
turning.However,
therotorandchambercanbesimplycleanedbydisassemblyand
washing.
3 Installationattitudecanbedifferentfromoriginalcalibration.Themetersarenormally
calibratedinahorizontallinewiththemeterinstalledupright.Ifadifferentorientation
isadvisedoruseditwillbenotedonthecalibrationcertificate.
Therewillbesomesmall
functionaldifferencesifthemeterisins talledon verticallyupwardfloworupsidedown.
LitreMeterdonotrecommendthemeterisi nstalledforverticallydownwardflow.
Performanceattheverylowestflowsmaybecompromisedunlesscalibratedinthe
correctorientation.
Anerroranalysis
systemisdescribedindocumentLM0638,VFFFaultFinding.
On-Site Flowmeter Calibration
ThisproceduremustbeintegratedwiththeapprovedHealthandSafetypolicyatsite.
Thepurposeofthissectionistooutlineageneralcalibrationprocedureforvariousflowmeters.Itwillmakespecific
referencetovolumetricflowmetersbutmostoftheprinciplesareequallyapplicabletootherflowmeasurement
technologies.Theimportantconsiderationisthatthereisaccesstothe readingoftheMeterUnderTest(MUT)anda
sufficientlyaccurateunderstandingoftheactualflowrate.
On-Site calibration
Theaimofthiscalibrationtechniqueistoprovidedatatotheusersuch
thattheinstrumentationcanbeadjustedtoshowthecorrectflowrate
and/ortotal
Thiswillbedoneby
o maintainingaknownflowrateand
o monitoringtheoutputoftheMUT
Precautions:
Flush/Cleanpipingpriortoinstallingthemeterorany
pipeworkchanges
Ensurethataliquidsystemhasnoairorvapour
entrained
Ensurethattheflowvariationsduringtestingareata
minimum,usuallybetterthan±1%variation
Ensuretherearenoprotrudinggasketsupstreamofthe
meter(forturbinemetersetc.)
Measurethepipingtoensurethatdimensionsand
finisharewithinstandardlimits
Ensurethatthemeasurementequipmentisinstalledin
accordancewithstandardsortothemanufacturer’s
recommendations
ThisishowLitreMetercalibrateflowmetersin
thefactory:
CALIBRATION
MostLitreMeterflowmetersarecalibrated
usingagravimetricmethod.Aflowrateis
establishedthroughtheflowmeterwiththe
outputflowreturningtothemainreservoir.At
thecommencementoftest,theoutputflowis
divertedintoaweightank.Whensufficient
volumeoffluidhasbeencollectedinrespectof
thatflowrate,theoutputflowisdivertedto
themainreservoironcemore.Thetimefor
thevolumeoffluidtobecollectedisrecorded,
togetherwiththenumberofpulsesproduced
bythetransmitter.Thedensityofthefluidis
determinedattimeofcalibration.Volume
dividedbytimeequalsflowrate.Thenumber
ofpulsesdividedbyvolumeequalsthepulses
perlitreMeterFactor.Thecalibration
certificateispreparedfromatableofthese
values.Ifonsitecalibrationisrequiredanda
knownvolumeoffluidflowcanbeestablished,
thenthesamecalculationsapply toreproduce
thecalibrationcertificate.