LMS COOLED INCUBATORS
Quick Start Instructions
Commissioning
• Remove all packing materials (CARE: door keys wrapped in shelf packaging).
• Check for signs of external damage (this must be reported to the supplier
immediately).
• Check supply voltage and cycles to ensure it is in accordance with your supply.
• Before connecting to the mains, please allow the Incubator to stand for a
minimum period of 24 hours to allow the refrigerant to settle after transportation
(CARE: failure to comply with this instruction may cause damage to this
instrument and invalidate the guarantee).
• After 24 hours you may switch the incubator on and start your application.
• Please ensure refrigeration switch on the rear of the unit is switched on.
• Maintain a minimum 100mm air gap around the incubator and ensure it is not
situated in direct sunlight.
Temperature Control
• The digital controller has been factory tuned for optimum performance and set at
20ºC.
To obtain the required temperature:
Press and hold the * button, this will display the set point (with the
programmable temperature controller the set point is the lower of the two
displays in orange) – press either the ? to increase or ? to decrease. When
the desired set point is reached release the * button. The incubator will now
start to react to your new set point temperature.
Temperature Controller with Programming
It is necessary to write a program before a program can be run. There are no stored
programs within the temperature controller at delivery unless this has been specified
otherwise. See general programming information Section 5