LogTag TRIX-8 V2 Work Instruction Manual Revision 10.17 Page 31/44
2.11 Temperature Cycle testing
Reason for test stage:
To subject LogTag products to temperature cycle stress.
Equipment required:-
Chest Freezer capable of achieving temperatures of -15ºC or lower.
rack or similar to hold LogTags in place with air space between each LogTag to allow flow of air for
cooling & warming.
Setup:-
Allow Freezer to reach stable temperature for 6 hours after switch on.
Procedure
Place logtags into freezer for 12 hours
Take logtags out of freezer & into room temp for 8 hours
Place logtags into freezer for 12 hours
Take logtags out of freezer & into room temp for 8 hours
Next production step