The raised side rail is forbidden to use in a bed, where fixing brackets’
hook-part remain free despite correct installation (danger of
disengaging)
Use of the raised side rail is prohibited if installation cannot be done
according to installation instructions (possible issue with suitability
between the raised side rail and beds’ side rail). Contact retailer in such
cases.
Using of thick and/or soft mattress increases suffocation risk between
mattresses' edge and side rail. Manufacturer cannot guarantee safety of
all mattress -types, user must estimate the safety of used mattress
based on mattresses' edges rigidity. Especially when air mattress is
used, these risks must be considered, and risk evaluation must be
conducted.
When using air mattress, make sure the mattress is set right way
around (head end of the mattress placed on head end of the bed)
The mattress must meet the bed manufacturer's requirements for
height and width.
7.6 IV pole (60121 or 60122)
The IV pole is used as a stand for infusions and blood bags when intravenous
fluids are administered. The IV pole is installed in a sleeve at the head end of
the nursing bed. The IV pole is secured by twisting it in the holder. The
maximum load (SWL) for the IV pole is 2 kg.
Image 44. IV pole in head end of the bed
Ensure that there is enough space above the bed. Keep in mind that
raising the bed also lifts the IV pole.