NORSK
Beregning
Kombinert klor = totalt klor - fritt klor
Tekniske data
Fritt klor: 0,05–6,0 mg/l
Cl
2
Totalt klor: 0,05–6,0 mg/l
Cl
2
(Nøyaktighet: 0-1 mg/l ± 0.2; 1-2 mg/l ±0.3; 2-3 mg/l ±0.4; 3-6 mg/l ±0.5)
Brome:
0,1 - 13,5 mg/l Br
(Nøyaktighet: 0-2 mg/l ± 0.5; 2-4 mg/l ±0.7; 4-7 mg/l ±0.9; 7-13 mg/l ±1.2mg/l)
pH-verdi: 6.5 - 8.4 pH
(Nøyaktighet ±0.2 pH)
Cyanursyre 1 -80 mg/l Cys
(Nøyaktighet ±10 mg/l)
Alkalinitet:
10 - 500 mg/l CaCO
3
(Nøyaktighet ±50 mg/l)
Lyskilde: LED, filter (
λ
= 528 nm)
Batteri: 9 V blokk-batteri (levetid 600 tester)
Autout-
kobling: Kobles automatisk av 5 minutter etter siste tastetrykk
Omgivelses- 5 -40° C,
forhold: 30-90% rel. fuktighet (ikke-kondenserende)
CE: DIN EN 55 022, 61 000-4-2, 61 000-4-8, 50 082-2,
50 081-1, DINVENV 50 140, 50204
Vedlikehold
1. Sørg for at instrumentet holdes rent.
2. Ikke bruk løsningsmidler eller slipemidler til å rengjøre instrumentet.
3. Tørk instrumentet når det ikke skal brukes, og oppbevar det ved
§§romtemperatur.
4. Fjern batteriet dersom instrumentet ikke brukes over en lengre periode..
Feilsøking
1. Bruk kun tabletter i hvit pose med blågrå skrift.
2. Ikke ta på tablettene, det kan føre til at de forurenses.
3. Lukk det grå lokket toppen av prøvekammeret før du
nuller ut / leser av.
4. Forsikre deg alltid om at prøven og lokket er skylt skikkelig
når du bytter fra en prøve til en annen.
• Sette i batteriet (9 volt, 6LR61)
BAT Bytt batteriene når denne meldingen vises i displayet.
Pass den svarte o-ringen når batteridekselet åpnes og lukkes!
6
7
9
TEST
water
tight
position
vann-
tett
posisjon