Models 375 (Including Series One), 375/1, 375/2 & 375/4 Section 5
Ludlum Measurements, Inc. Page 5-5 May 2017
Ethernet Interface Option
A 10-BaseT Ethernet interface may be added internally for network
reporting, using Ludlum software:
4558-098 LMI “Ethernet” Hardware Interface
4558-105 LMI “Webpage” Hardware Interface
Either the Ethernet software (1370-055) or the Webpage software (1370-
077) must be purchased separately (site-licensed).
Time and Date Stamp Option
Description:
When an alarm or failure occurs, the Model 375 will print the current
reading, date, time, and either ALARM or FAIL to the RS-232 port. The Model
375 will print once every 30 seconds as long as the alarm or fail condition is
present.
Setup:
You will need the following: a Model 375 instrument, a CBM-910 40-
column printer, and a cable (8558-142).
The printer should be configured at 2400 bps (baud), no parity, 8 data
bits, 1 stop bit, and no handshaking. See printer manual for proper setup
instructions.
Setting the date and time:
Check the month and day (MMDD) by pressing the
LOW ALARM and
HIGH ALARM buttons simultaneously. The month and day will be
displayed as long as those buttons are pressed. The month and day can be
set from 0101 to 1231.
Check the year (YYYY) by pressing the LOW ALARM and CAL CONST
buttons simultaneously. The year will be displayed as long as those
buttons are pressed. The year can be adjusted from 0000 to 9999.
Check the hours and minutes (HHMM) by pressing the
LOW ALARM and
DEAD TIME buttons simultaneously. The hours and minutes will be
displayed as long as those buttons are pressed. The hours and minutes
can be adjusted from 0000 to 2359.