EasyManua.ls Logo

Lutron Electronics SOIL PH METER - 4 Ph CALIBRATING PROCEDURE; Calibrating Consideration; Requiring Equipment for Calibration; Calibration Procedures

Lutron Electronics SOIL PH METER
12 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
4. pH CALIBRATING PROCEDURE
4-1 Calibrating Consideration
The pH meter already calibrated by mV signal that
simulated from the ideal pH ELECTRODE mV output
( base on 25 environment ). However due to (a) An
ideal electrode will produce 0 mV at pH 7.00, but most
electrodes are slightly off. (b) The measuring environment
temperature may not near 25 ( 15 to 35 ). ℃℃
If intend to keep instrument combined electrode within
high accuracy, it is necessary to make the following
calibration procedures ( 4-3 ).
4-2 Requiring Equipment for Calibration
1)Combination soil pH electrode ( PE-06HD, included ).
2)Buffer solutions :
pH 7.00 buffer solution ( PH-07A, included ).
pH 4.00 buffer solution ( PH-04A, included ).
4-3 Calibration procedures
pH 7 calibration
1)Connect the pH electrode plug ( 3-10, Fig. 1 ) to the BNC
socket ( 3-5, Fig. 1 ) and place electrode into pH 7.00
buffer solution ( PH-07A ).
2)Power on the instrument by pressing the " Power button "
( 3-2, Fig. 1 ).
3)Press the " Hold button " ( 3-3, Fig. 1 ) once.
the display will show the " HOLD " symbol.
Press the " button " ( 3-3, Fig. 1 ) once, the display
will show following screen in sequence.
5

Related product manuals