Do you have a question about the Lutron Electronics SL-4010 and is the answer not in the manual?
| Brand | Lutron Electronics |
|---|---|
| Model | SL-4010 |
| Category | Measuring Instruments |
| Language | English |
Highlights include large LCD, 'A' frequency weighting, 'Fast' time weighting, built-in calibration VR, condenser microphone, data hold, and low battery indicator.
Details on the 3 1/2 digit LCD, measurement ranges (35-130 dB), resolution (0.1 dB), and frequency response (31.5 Hz-8,000 Hz).
Specifies 'A' frequency weighting network meeting IEC 61672 class 2, simulated as human ear response.
Details 'Fast' time weighting and accuracy specifications for various frequencies under 94 dB input signal.
Information on the built-in calibration VR, microphone type, data hold function, and optional calibrator.
Specifies operating temperature (0-50°C), humidity (<80% RH), power supply (9V DC), and consumption (approx. 6 mA).
Provides unit dimensions (250x70x28mm), weight (250g), and lists standard and optional accessories.
Identifies the electric condenser microphone and the digital display unit on the front panel.
Locates range switch, calibration VR, power/hold switch, and range indicators on the front panel.
Indicates the position of the battery compartment and cover on the front panel.
Guides on powering on, selecting the measuring range, taking readings, and using the data hold function.
Instructions for replacing the battery when the low battery indicator appears.
Details on preparing an optional sound calibrator and connecting it to the sound level meter.
Instructions to set the range switch and adjust the external calibration VR for accurate readings.
Presents a table of A-weighting characteristics and tolerances for various frequencies (31.5 Hz to 8 KHz).