6)
Data Record ( Max. and Min. reading )
The Data Record function displays the maximum and
minimum readings. To start the Data Record function,
press the " Record Button " ( 3-3, Fig. 1 ) once.
" REC " symbol will appear on the LCD display.
With the " REC " symbol on the display :
a)Push the " Recall Button " ( 3-4, Fig. 1 ) once, the
" Max " symbol along with the maximum value will
appear on the display.
b)Push the " Recall Button " again, the " Min " symbol along
with the minimum value will appear on the display.
c) To exit the memory record function, push the " Record
Button " once again. The display will revert back to the
current reading.
7)
Offset Value adjustment
@Caused by the environment temperature change or other
reasons.... The measuring value of the thermocouple
sensor input (type K/J/T/E/R) may exist the drift of small
digits.
@If find that the measuring values exist deviation of small
digits especially when measure the low temperature, then
can adjust the " Offset Adjust VR " to make the
compensation until get the correct reading exactly.
The offset adjustment procedures are following:
a)Plug in a type k thermocouple probe to the input
socket & Set the " Sensor Select Button " ( 3-5, Fig. 1 )
to the " Type K". Now the meter is measured the
environment Temperature value.