What does t04 Error Code mean on Macurco Gas Detectors?
- BBenjamin SteeleSep 9, 2025
A t04 Error Code on Macurco Gas Detectors indicates an EEPROM bad checksum.
What does t04 Error Code mean on Macurco Gas Detectors?
A t04 Error Code on Macurco Gas Detectors indicates an EEPROM bad checksum.
What does t40 Error Code mean on Macurco Gas Detectors?
A t40 Error Code on Macurco Gas Detectors indicates a Factory calibration failure.
What does t80 Error Code mean on Macurco Gas Detectors?
A t80 Error Code on Macurco Gas Detectors indicates Read ADC failure.
What does t100 Error Code mean on Macurco Gas Detectors?
A t100 Error Code on Macurco Gas Detectors indicates Under range.
Essential safety warnings and precautions for operating the refrigerant detector.
Details the intended applications and capabilities of the RD-xx refrigerant detector.
Specifies limitations and environments where the RD-xx should not be used.
Lists the key features and technical specifications of the RD-xx detector.
Guidance on selecting the optimal mounting location for the RD-xx detector.
Step-by-step instructions for installing the RD-xx unit, including wiring connections.
Details wiring connections for the 6-Series low voltage model, including power, fan, alarm, and 4-20mA signals.
Details wiring connections for the 12-Series line voltage model, including mains power and safety ground.
Describes the unit's self-test cycle upon initial power-up and normal operation startup.
Explains how the RD-xx display shows gas concentration and status when enabled.
Explains how the RD-xx display behaves when the display function is turned off.
Details the 4-20mA output loop behavior for clean air, gas readings, and trouble conditions.
Lists and explains the default factory settings for various RD-xx operational parameters.
Detailed instructions for configuring various RD-xx operational settings, including relays, tests, and calibration periods.
Overview of the RD-xx's self-diagnostic features and error reporting mechanisms.
Guidance on troubleshooting issues related to the 4-20mA signal output.
List and explanation of various error codes ('tXX') indicating detector malfunctions.
Information on substances that can affect sensor performance or cause nuisance alarms.
Describes the end-of-life signal for the sensor and how to manage it.
Procedure for resetting the sensor life counter after replacement or maintenance.
Instructions for cleaning the external surfaces of the RD-xx detector.
Steps to perform a basic operational test of the RD-xx unit's indicator light and relays.
Allows manual initiation of individual tests for relays, analog output, and sensor response.
Information on calibration procedures and the necessary test kits for the RD-xx.
Detailed steps for performing gas tests, including fan relay, alarm relay, and 4-20mA loop verification.
Step-by-step guide for performing field calibration of the RD-xx detector.