EasyManua.ls Logo

Maico MI 24 - Page 50

Maico MI 24
71 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation Manual
touchTymp MI 24
and
MI 34
Version
8107420 Rev. 6
03/09/2018
5.6.2 Settings Basic General
Figure 87
Start Screen: Adjust the start screen to your needs.
Choose Tympanometry or Tympanometry and
Acoustic Reflexes for upon boot-up the chosen screen is
automatically entered (Figure 87).
NOTE: MI 34 Version will offer Reflex Decay and ETF
within this menu setting.
Figure 88
Probe out of holder: Select Probe out of holder for
automatic changing from the setting or home screen to the
test screen as soon as the probe is taken out of the probe
holder (Figure 88). Setting is inactive if Audiometry is
selected as Start Screen (Figure 87).
Figure 89
Language: Choose one of the supported languages
incorporated in the device (Figure 89).
Figure 90
Standby: Set the period of inactivity, after which the
display will turn off. Pressing the screen or the Front key
will awaken the device (Figure 90).
NOTE: It is possible to turn off this function by setting
the value to never”. When in standby mode the probe
light is lit to indicate device is on.
Figure 91
Auto shutdown: After a period of inactivity (greater than
the Standby Mode setting) the device will turn off
automatically (Figure 91).
NOTE: Data will be lost when the device turns off. It is
possible to turn off this function by setting the value to
never”.
Figure 92
Brightness: Set the maximum brightness of the display
(Figure 92).

Other manuals for Maico MI 24