manual stirrer operation
mode
0.0 – continuous work
>0.0 – stirrer operation for a specified time in
the range of from 1 min. to 60 min., every 1
min.
delay of registering of the
maximum / minimum
values of temperature and
delay of activation of the
temperature alarm
setting the time from 0 h to 4 h, every 0,1 h
lower temperature of the
alarm
setting the temperature from -40
O
C to
+120
O
C, every 1.0
O
C
upper temperature of the
alarm
setting the temperature from -40
O
C to
+120
O
C, every 1.0
O
C
visual and sound alarm
of exceeding the set
temperature range
(parameters A1 and A2)
scaling the control
sensor**
setting the operation
of the limit switch
of opening the tank’s flap
0 – closing the limit switch: raising the flap
1 – opening the limit switch: raising the flap
delay of switching on the
chiller (unit) and the stirrer
after connecting the
regulator to the mains
or when power supply
decay occurs
setting the time from 1 sec. to 999 sec., every
1 sec.
display status when
the control is switched off
0 – the dot is displayed
1 – the currently measured temperature is
displayed
return to factory settings
0 – no return to the factory settings
1 – after setting parameter to 1 and turning
off the regulator from the power supply,
switching it on again restores the factory
settings and the parameter value
automatically goes back to 0
** set at the manufacturing stage (do not change without obvious need)