Do you have a question about the Mallinckrodt INOmax DS and is the answer not in the manual?
Avoid recirculation; can lead to higher NO2, altered NO concentrations, and reduced O2.
Sterilize Injector Module between patient uses if in wet/humidified breathing circuit.
Ensure flow out of absorber passes through Injector Module in the direction of the arrow.
N2O affects Set NO vs. measured NO; 50% N2O/50% O2 composition shows ~7% less measured NO.
2% v/v isoflurane results in ~3% higher measured NO value compared to 100% O2.
Sudden changes in anesthetic agent concentration can cause brief transient changes in measured NO/NO2 values.
System performs as specified with fresh gas flow rates equal to or greater than patient minute volume.
Circuit between sample tee and patient Y should be 6-12 inches for accurate gas sampling.
For OR systems, place Injector Module upstream of inspiratory flow sensor.
| Device Type | Nitric Oxide Delivery System |
|---|---|
| Monitoring Capabilities | Continuous monitoring of NO, NO2, and O2 concentrations |
| Power Supply | 100-240 VAC, 50/60 Hz |
| Brand Name | INOmax |
| Manufacturer | Mallinckrodt |
| Indication | Treatment of term and near-term infants with hypoxic respiratory failure. |
| Gas Source | Nitric Oxide (NO) |
| Concentration Range | 0-80 ppm |
| Flow Rate | 0.05 to 10 L/min |
| Alarms | High NO2, Low NO, High NO, Low O2, System Malfunction |
Loading...