Part No. 20003 Rev - 01
Part No. 20717 Rev-01
2014-07
Operation Manual
(800 ppm INOMAX® (nitric oxide) for inhalation)
Series 3 software

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Mallinckrodt INOMAX DSir Plus and is the answer not in the manual?

Summary

General Information

Indications for Use

Details the intended use and therapeutic application of the INOmax DSIR system for nitric oxide delivery.

Automated Pre-Use Checkout

High Pressure Leak Test and Automated Purge

Procedures for performing a leak test and automated purge of the system before patient use.

Performance Test

Details the steps for conducting the performance test to verify system accuracy.

Patient Application

Before Operation

Key operational considerations and warnings before connecting the INOmax DSIR to a patient.

Use Outside of Product Labeling

Restricts system use to approved applications, patient populations, and conditions.

Connection to Ventilator Breathing Circuit

Guidelines for connecting the INOmax DSIR to various ventilator breathing circuits.

INOblender Operation

Instructions for operating the INOblender, including warnings for specific use cases.

Integrated Pneumatic Backup NO Delivery

Explains the function and activation of the integrated pneumatic backup for NO delivery.

Transport

Transport Options

Overview of different transport scenarios and recommendations for equipment.

InterHospital Transport

Recommendations and equipment list for inter-hospital transport of INOmax DSIR system.

Transport Regulator;Cap Assembly Application

Detailed instructions for applying the transport regulator/cap assembly.

Alarms

High Priority Alarms

Lists high priority alarms, their possible causes, and recommended actions.

Calibration

Low Calibration

Procedure for performing automatic or manual low calibration using room air.

Oxygen Sensor High Calibration

Steps for performing high calibration of the O2 sensor using 100% oxygen.

NO Sensor High Calibration

Procedure for high calibration of the NO sensor using calibration gas.

NO2 Sensor High Calibration

Procedure for high calibration of the NO2 sensor using calibration gas.

Maintenance

Replacing the O2, NO and NO2 Sensors

Detailed instructions for replacing the O2, NO, and NO2 sensors.

Cylinder Leak Check

Steps to check for leaks in INOMAX regulators and cylinders using soapy water.

Product Specifications

Ventilator Compatibility

Lists ventilators and breathing systems validated for use with the INOmax DSIR in the US.

NO Delivery Specifications

Details on NO delivery range, resolution, accuracy, pressure, and gas composition.

Appendix

Initial Connections

Step-by-step guide for connecting the injector module, cables, and gas lines.

High Pressure Leak Test

Procedure for performing a high pressure leak test on the INOMAX cylinder and regulator.

Manual Purge and Alarm Verification

Steps for manual purging and verifying system alarms during pre-use.

Integrated Pneumatic Backup INOMAX Delivery Test

Tests the integrated pneumatic backup delivery function for NO delivery.

INOblender Test

Procedure for testing the INOblender's functionality with the INOmax DSIR.

Mallinckrodt INOMAX DSir Plus Specifications