Standby mode (when
applicable)
KMP-017X
IN
=
On Standby Alarm
63 psi
STANDBY
• Connect the central air supply.
The compressor is automatically
switched off when the air pressure
exceeds 44 psi (300 kPa/3.0 bar) for
more than 30 seconds.
-
KMP-046X
IN
• Disconnect the central air supply
KMP-030X
=
On Standby Alarm
61 psi
12345 hours
• Decrease the internal pressure with the
manual drain valve. When the air pressure
drops below approximately 36 psi (250
kPa/2.5 bar) for more than 2 seconds the
compressor starts.
Checks during operation
KMP-018X
1. Check the water level in the drainage
bottle regularly.
2. Empty the bottle if necessary; the
compressor does not need to be turned
off.
KMP-019X
• To check the internal air reservoir for water,
press the pin on the drain valve. Check that
no water, or only a very small amount of
water flushes out through the drain valve.
14
COMPRESSOR MINI, User´s Manual
English version
28.0
4. Function check