Alarms
DescriptionAlarms
The red LED and the alarm buzzer are activated if the
temperature inside the compressor unit exceeds 158 °F
±5 °F (70 °C ±3 °C) [temp. difference]. The alarm is
activated until the temperature drops below 131 °F ±11
°F (55 °C ±6 °C) [temp. difference], or the compressor is
switched off.
High temp alarm 1
KMP-036US
62 psi
High temp 1
On Standby Alarm
The red LED and the alarm buzzer are activated if the
temperature difference between ambient air and the
temperature in the dryer is less than approx. 4 °F (2 °C)
after 15 minutes operating time. The alarm is activated
until the temperature difference between ambient air and
the temperature in the dryer exceeds approx. 7 °F (4 °C).
High temp alarm 2
KMP-037US
62 psi
High temp 2
On Standby Alarm
The red LED and the alarm buzzer are activated if the
temperature in the dryer exceeds 131 °F ±11 °F (55 °C
±6 °C) [temp. difference]. The alarm is activated until the
temperature inside the dryer drops below 122 °F ±11 °F
(50°C ±6 °C) [temp. difference].
High temp alarm 3
KMP-038US
62 psi
High temp 3
On Standby Alarm
The red LED and alarm buzzer are activated if the
compressed air pressure drops and stays below 29 psi
(200 kPa/2 bar) for more than 5 seconds. The alarm is
activated until the air pressure rises above approx. 36 psi
(250 kPa/2.5 bar) for more than 5 seconds, or the
compressor is switched off.
Low pressure alarm
KMP-043US
24 psi
Low pressure
On Standby Alarm
The red LED and alarm buzzer are activated if the
compressed air pressure rises and stays above 94 psi
(650 kPa/6.5 bar) for more than 0.5 second. The alarm is
active until the air pressure drops and stays below 87 psi
(600 kPa/6 bar) for more than 5 seconds, or the
compressor is switched off.
High Pressure alarm
KMP-044US
100 psi
High pressure
On Standby Alarm
Power failure. The alarm buzzer is activated for about 12
seconds.
Power failure alarm
COMPRESSOR MINI, User´s Manual
English version
28.0
19
7. Alarms