MAQUET FLOW-I COVID-19 QUICKGUIDE
TWO RULES OF FLOW-I
1. Make sure it is switched on
2. READ THE SCREEN
FLOW-I
Screen with touch
display area for
ventilatory functions
Rotary knob for selecting, changing
and confirming values on the screen.
Membrane buttons for quick access
to special windows on screen eg
trends, silence alarms etc
HEPA filter
CO2
absorber
(sodalime)
Power button
Emergency ventilation allows
patient to be manually ventilated in
the case of system failure. When the
emergency switch is set to ‘ON’ the
anaesthetic system will shut down.
Fresh gas flow is replaced by 100%
O2 and a manual APL valve under the
flap
Water trap
needs to be
changed
when half
full (or every
17 hours
even if it
isn’t half
full)

Other manuals for Maquet FLOW-I

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Maquet FLOW-I and is the answer not in the manual?

Maquet FLOW-I Specifications

General IconGeneral
TypeAnesthesia Machine
ModelFLOW-I
Power Supply100-240 V AC, 50/60 Hz
ManufacturerMaquet
Ventilation ModesPressure Support
Display15-inch touchscreen display
Integrated FeaturesIntegrated ventilator, breathing system, monitoring
Oxygen MonitoringIntegrated oxygen monitoring
Vaporizer OptionsIsoflurane, Sevoflurane, Desflurane
Gas SupplyO2, N2O, Air