EasyManua.ls Logo

Marconi Instruments TF 2700 - Reading the Result

Marconi Instruments TF 2700
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
2..1.2. READING THE RESULT
The main BALANCE controls of the bridge comprise a ten position
switch with a variable interpoIating contraI mounted concentrically. The
switch provides increments of O. l between each position. The variable
contraI covers one incremento with slight overIap; its calibration of 00
to 100 giving the second decimaI piace withintermediate divisions of 0.002..
The overall calibration is therefore O to aver 1.1; examples of BALANCE
contraI readings are given in Fig. 2.. l.
( (
c.}o.s.. (o) 00604
QJ( (
(c) ,- (o) 0-.0.
Fig. 2.. l BaIance ContraI Indication
The table of range full scale values is linked to the BALANCE controls
by the RANGE MULTIPLIER switch. The correct impedance value is obt-
ained by multiplying the BALANCE contraI reading by the fun scale value
indicated by the FUNCTION and RANGE MULTIPLIER switches. Fig. 2.2
shows the bridge set to a value of 66.6 mH.
...
~
""K
Fig. 2.2
Test Component Value - 66.6 mH
8