1.2 DATA SUMMARY
RANGES
C apacitance: O. 5 pF to 1100 I1F in 8 range s of 110 pF to
1100 I1F full scale.
Inductance: O. 2 I1H to 110 H in 8 range s of Il I1H to
110 H full scale.
Resistance: lO ni1 to Il M1 in 8 ranges of 1.1 a to
Il MrZ full scale.
Q lndication: O to lO at l kc/s.
D Indication: O to 0.1 or O to lO at l kc/s.
ACCURACY
Basic Accuracy: :1:1% of reading :1:0. 1"/0 of range full scale.
Inductance: Additional -0.2"/0 of reading. +0 -0.3 I1H.
First Ranges only:
Capacitance. 100 I1F
201 f d .
:1:1 Fto 1100 I1F: :I: IO o rea lng. jJ...
Inductance. 0.2jJ.H +2"/0 -10"/0 of reading. +0 -O. 3jJ.H dependent
to lljJ.H: upon Q.
Resistance. lO ~
201
f di :1:1 -..n
l :I: IO o rea ng. LLA'.
to l. a:
Residuals to be COI approx. O.l pF.
subtracted: Lo. approx. 0.2 jJ.H.
Ro. approx. 2~.
BRIDGE SOURCE
InternaI: l kc/s :1:50;0 oscillator far a.c. measurements
of C. L, and R. 9 V battery for d.c. me a-
surements of R between 1.2 a and 50 ~.