RTD Type Range (°C) Accuracy (°C)
Minimum Maximum Measure 4W/Source Measure 3W
300.0 400.0 0.5 0.7
400.0 630.0 0.5 0.8
630.0 800.0 0.6 0.9
PT385, 100 ohm -200.0 -80.0 0.1 0.3
-80.0 0.0 0.2 0.4
0.0 100.0 0.2 0.4
100.0 300.0 0.2 0.4
300.0 400.0 0.3 0.5
400.0 630.0 0.3 0.5
630.0 800.0 0.4 0.6
PT3926, 100 ohm -200.0 -80.0 0.1 0.2
-80.0 0.0 0.2 0.3
0.0 100.0 0.2 0.3
100.0 300.0 0.2 0.3
300.0 400.0 0.3 0.4
400.0 630.0 0.3 0.5
PT3916, 100 ohm -200.0 -80.0 0.1 0.2
-80.0 0.0 0.2 0.3
0.0 100.0 0.2 0.3
100.0 260.0 0.2 0.3
260.0 300.0 0.3 0.4
300.0 400.0 0.3 0.4
400.0 630.0 0.3 0.5
PT385, 200 ohm -200.0 -80.0 0.6 0.6
-80.0 0.0 0.7 0.7
0.0 100.0 0.7 0.8
100.0 260.0 0.7 0.8
260.0 300.0 0.7 0.8
300.0 400.0 0.9 0.9
400.0 630.0 0.9 1.0
PT385, 500 ohm -200.0 -80.0 0.2 0.3
-80.0 0.0 0.3 0.3
0.0 100.0 0.3 0.3
100.0 260.0 0.4 0.4
260.0 300.0 0.4 0.4
300.0 400.0 0.4 0.5
400.0 630.0 0.5 0.5
52
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com