PERFORMANCE SPECIFICATIONS
Table information provides
Arms values measured using the DBI sensors with Masimo SET Oximetry
Technology in a clinical study.
MEASURED ARMS VALUES
RANGE
Arms
90–100% 1.01%
80–90% 1.54%
70–80% 2.06%
OVERALL CLAIMED ACCURACY VALUE
70–100% 1.60%
SaO2 versus error (SpO2 - SaO2) with linear regression fit and upper 95% and lower 95% limits of agreement.