Maxim Integrated Page 10 of 15
Table 5. Format of Received Samples—SpO
2
Calibration Mode
*If the output mode includes the sensor.
**If the output mode includes the algorithm.
DATA SOURCE BYTE INDEX DATA ITEM
DESCRIPTION
MAX30101
(12 Bytes)*
Accelerometer
(6 Bytes)*
12 accelX 2
Two's complement. LSB =
0.001g
14 accelY 2
Two's complement. LSB =
0.001g
16 accelZ 2
Two's complement. LSB =
0.001g
HR/SpO
2
Algorithm
(12 Bytes)**
20
1
Calculated confidence level
in %
23
Algorithm
state
1
0: No object is detected
1: Something is on sensor
2: Another object is detected
26
Algorithm
status
1
Algorithm current status:
0: Success
1: Not ready
-1: Something is on sensor
-2: Device excessive motion
-3: No object
-4: Pressing too hard
-5: Object instead of finger
-6: Finger excessive motion