Doc no: PT00098114 Revision A Aircraft Medical Ltd © 2019 www.medtronic.com/mac-first
11
3 Cleaning and Low Level Disinfection
The device and battery should be cleaned and low level disinfected separately after each patient use.
Cleaning and low level disinfection should be carried out as per section 3.1, 3.2, 3.3, and 3.4 of the
manual in accordance with local regulations. Follow hospital policy and protocol when handling and
cleaning soiled items.
The instructions provided have been validated by Aircraft Medical as being capable of preparing the
device for re-use. It remains the responsibility of the processor to ensure that the reprocessing as
actually performed using equipment, materials and personnel in the reprocessing facility achieves the
desired result. This normally requires validation and routine monitoring of the process.
The following methods have been approved by Aircraft Medical:
PRODUCT CLEANING
LOW LEVEL
DISINFECTION
McGRATH®
Disposable
Laryngoscope Blade
Supplied Sterile
Do Not
Reprocess
Supplied Sterile
Do Not
Reprocess
McGRATH® MAC
with McGRATH®
3.6V Battery removed
70% IPA* Wipe 70% IPA Wipe
McGRATH®
3.6V Battery
70% IPA Wipe 70% IPA Wipe
*IPA = Isopropyl Alcohol