2.3. Assessment
Every EPIRB returned for servicing, of whatever nature, must be fully assessed to
determine its operational status. The diagram shows the tasks which must be carried out
for each servicing situation.
If the EPIRB is opened for any reason, it is recommended a new sealing kit is fitted
.
Assessment form
The assessment form, shown on the next page, is an essential tool in performing
diagnosis. It may be used to support customer billing, and must accompany any warranty
return to the manufacturer.
The first section of the form records the customer, the EPIRB and the reason for return
(as 'Customer comments'). It is essential to record the identity details from the EPIRB
rear label, as these will be necessary for comparison with the transmitted data. The
expiry date must be noted; if less than 6 months life remains, a battery change should be
recommended to the customer.
Opening the EPIRB for internal inspection is only required if a problem is found.
RECEIPT