page 26
PORT - Communicates with peripheral device.
Transmission of the displayed load reading
can be set to include unit of measurement
(UNITS ON or OFF) and BAUD rate can
also be set.
STORE MEM - Stores a single load reading to the internal
memory. With this option selected, pressing
the TXD key when in the MAX modes of the
main display will send the displayed value to
memory. Up to 100 readings may be stored
in the memory. As each reading is stored, a
‘REC’ symbol appears on the main display
to indicate a reading has been stored in the
memory.
SEND MEM - Sends all load readings stored in the internal
memory to a peripheral device (e.g PC or
data logger).
CLEAR MEM – Erases all load readings stored in memory.
When setting PORT you will now access sub-menu 1.
Transmission of the displayed load reading can be set to
include the unit of measurement (UNITS ON or OFF).
Use the UP or DOWN key to position the arrow cursor at
either UNITS OFF or ON. Press ENTER key to select.
Sign off or on can now be set. This will transmit a negative
sign for compression readings if set to ON.
Use the UP or DOWN key to position the arrow cursor at
either SIGN OFF or ON.
Press ENTER key to select.
The transmission (or Baud) rate can now be set.
Use the UP or DOWN key to position the arrow cursor at the
relevant speed (9600, 19200, 57600 or 115200).
Press ENTER key to select.
Port sub-menu 1
Port sub-menu 2
Port sub-menu 3