Electromagnetic immunity
The electric breastpump Swing is intended for use in the electromagnetic environment
specied below. The customer or the user of the electric breastpump Swing should assure
that it is used in such an environment.
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic
environment
– guidance
Electrostatic
discharge ( ESD )
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood,
concrete or ceramic tile.
If floors are covered with
synthetic material, the
relative humidity should
be at least 30%.
Electrical fast
transient / burst
IEC 61000-4-4
± 2 kV for power
supply lines
± 1 kV for input /
output lines
± 2 kV for power
supply lines
Not applicable
a
Mains power quality
should be that of a
typical commercial or
hospital environment.
Surge
IEC 61000-4-5
± 1 kV line( s ) to
line(s)
± 2 kV line( s ) to
earth
± 1 kV
differential
mode
Not applicable
b
Mains power quality
should be that of a
typical commercial or
hospital environment.
a
No input/output lines
b
No earth, class II equipment
Electromagnetic Compatibility (EMC, IEC 60601-1-2:2007, Table 2)