User Manual

Questions and Answers

  • J
    Jennifer KellyAug 13, 2025
    How to fix no display on Medisafe SI DIGITAL Medical Equipment?
    • J
      jasonthomasAug 13, 2025
      If the front panel of your Medisafe Medical Equipment isn't displaying anything, first, ensure the unit is properly connected to a working power outlet. Next, inspect the fuses located at the rear of the unit to see if they are intact. For continued safe operation, always replace fuses with the same type and rating. If problems persist, it is recommended to consult a qualified electrician or your Medisafe distributor.
  • N
    Natalie WatkinsAug 16, 2025
    What to do if Medisafe Medical Equipment fails to operate and shows 'LU'?
    • T
      Traci CampbellAug 16, 2025
      If your Medisafe Medical Equipment is showing “LU” in temp mode and fails to operate, make sure the lid is completely closed and that the instruments inside aren't preventing it from doing so. Also, check that the magnet is in place on the top left of the lid, under the seal.
  • A
    Amanda FergusonAug 20, 2025
    Why is my Medisafe SI DIGITAL Medical Equipment performance reduced?
    • L
      Lucas PhillipsAug 20, 2025
      If your Medisafe Medical Equipment is experiencing reduced performance, verify that you are using a genuine Medisafe basket and that the basket feet are correctly positioned, making contact with the bottom of the tank. If it's the first cycle, run a 15-minute “de-gas” cycle. Also, confirm that you're using genuine Medisafe 3E-zyme (non-foam) wash chemical, and ensure instruments are not excessively soiled and are cleaned within 2 hours of use.
  • J
    johnharrisAug 25, 2025
    What to do if Medisafe SI DIGITAL Medical Equipment has no irrigation?
    • T
      Tracy ParsonsAug 25, 2025
      If your Medisafe Medical Equipment isn't irrigating, check the drain filter for any blockages and replace it if necessary. Also, inspect the irrigation system for obstructions by using a soiled hollow test instrument, and then run a cycle to evaluate the cleaning efficiency.
  • D
    dixonanthonyAug 27, 2025
    How to troubleshoot no ultrasonic activity in Medisafe SI DIGITAL Medical Equipment?
    • D
      derrick69Aug 28, 2025
      If your Medisafe Medical Equipment shows no ultrasonic activity, first ensure that the unit is switched on and in operational mode. Next, check that the lid is closed properly if an interlock is fitted. Lastly, verify that the water level is at the fill line or at least reaching the water sensor.
  • J
    Jeffrey CooperAug 31, 2025
    Why does Medisafe SI DIGITAL tank require too much water?
    • J
      Jason EllisAug 31, 2025
      If the tank of your Medisafe Medical Equipment requires excessive water to reach the full mark, check that the drain valve is not being held open by debris. It is recommended to seek professional advice or turn off the unit during inspection.
  • G
    Gloria BentonSep 4, 2025
    What causes water to leak from the lid of Medisafe SI DIGITAL?
    • S
      Sharon WardSep 4, 2025
      If water is leaking from the lid area of your Medisafe Medical Equipment, first check that the instruments are laying flat in the basket and not pointing upwards towards the lid. Also, inspect the lid seal for any damage.
  • K
    kdrakeSep 7, 2025
    What to do if Medisafe Medical Equipment fails to drain?
    • I
      ihollowaySep 8, 2025
      If your Medisafe Medical Equipment fails to drain, check the drain pipe for any blockages.

Summary

Operation

Memorandum of Registration

Details for product registration and warranty, including owner and installation information.

Important Information

Key safety warnings, electrical details, intended purpose, heat, decontamination, and manufacturer information.

Introduction

Overview of the instrument cleaner's intended use, performance, and declaration of conformity.

Identifying Controls

Explains the control panel layout, buttons, and LED indicators for the instrument cleaner.

Operating Instructions

Guide on loading instruments into the basket for a wash cycle, with important precautions.

Operating a Wash Cycle

Step-by-step instructions for running a wash cycle, including pre-cycle checks and pausing.

Advanced Functions

Instructions for disabling heaters and changing default wash time and temperature settings.

Routine Checks: Pre-Refill and Daily Procedures

Pre-refill and daily maintenance checks for the instrument cleaner.

Routine Checks: Weekly Procedures and Fuse Replacement

Weekly maintenance checks and fuse replacement procedure for the instrument cleaner.

Product Warranty

Details of the one-year warranty, basic warranty items, and specific excluded items.

Installation

Unpacking

Guide to unpacking the instrument cleaner, checking enclosures, and ordering replacement parts.

Technical Specification

Detailed technical specifications including dimensions, power, water supply, and chemical data.

Chemical Data

Safety and handling information for the 3E-Zyme cleaning fluid.

Dimensions

Illustrates the outer cabinet and wash chamber dimensions, and required installation clearances.

Installation Procedure

Step-by-step instructions for connecting the instrument cleaner to power, water, and drain.

Operating Process

Overview of the wash cycle, pre-cycle checks, and the functioning of the CPE pump.

Troubleshooting Guide

Basic troubleshooting steps for common operational errors and issues.

Service Record

Template for recording service history and scheduling future maintenance.

Decontamination Certificate

Form to certify that the machine has been disinfected before return or service.

Medisafe SI DIGITAL Specifications

General IconGeneral
TypeSyringe Infusion Pump
Infusion Rate Range0.1 - 1200 ml/h (depending on syringe size)
Volume Range0.1 mL to 999.9 mL
Power Supply100-240V AC, 50/60 Hz
DisplayLCD
AlarmsOcclusion, Low Battery
Dimensions120 mm x 120 mm
Syringe Compatibility10ml, 20ml, 30ml, 50/60ml