Troubleshooting
Use this basic guide to overcome any operational errors which may be encountered whilst using
your Medisafe SI Digital Cannulated Instrument Cleaner. If the problem remains unresolved after
consulting this guide then contact your supplier.
Section 2-7 - Troubleshooting Guide User Manual
Ref: UM SI Digital T 22 07/08 Rev.A
Problem Possible cause(s) and Solution(s)
No display on front
panel
Check that unit is connected to a live mains supply
Check that the fuses tted in the fuse-holders located at the rear of
the unit are intact
Note: for continued safe operation, always replace fuses with
the same type and rating. If in doubt, consult a suitably qualied
electrician or your Medisafe distributor
No irrigation Check that drain lter is not blocked - replace if necessary
Check for obstruction in the irrigation system by using a soiled hollow
test instrument, run cycle to check cleaning efciency
No ultrasonic
activity
Check that unit is switched on and in operational mode
Check lid is closed if interlock tted.
Check water is up to ll level or a least to water sensor.
Reduced
performance
Check that a genuine Medisafe basket is being used
Check that basket feet are in place and making contact with bottom
of tank.
If it is the rst cycle of the day then run a 15 minute “de-gas” cycle
to remove any dissolved gasses
Check that genuine Medisafe 3E-zyme (non-foam) wash chemical is
being used
Ensure that instruments are not excessively soiled
Ensure that instruments are being cleaned within 2 hours of use
Tank requires
excessive water to
reach full mark
Check that drain valve is not being held open by debris (seek advice
or turn off unit)