TREADMILLS MOD. 870A/870S/870C
Rev.1/01.07.2019
© Medisoft RAM Italia S.r.l.. All rights reserved.
26
4 Patient security
4.1 EMERGENCY STOP BUTTONS
The treadmill Medisoft RAM is the only device equipped with two emergency
stop buttons positioned frontally to the patient on the support structure ( when
the patient is on walking surface ) to the right and left sides . Are highly visible ,
fast access and usable.
By pressing the emergency button the treadmill as it turns off because the
power supply is disabled.
The running belt will stop by inertia and it will take about 5-8 seconds depending
on the speed that the treadmill had when the button was pressed.
There are informational stickers on the buttons that alert you to wait about thirty
seconds before reactivating the button , using a small rotating counterclockwise
so the treadmill will come out of safety status and back to work again.
If you do not wait 30 seconds the speed will not work .