8. Insert the biopsy needle into the guide tube and make sure that the biopsy bracket is
firmly attached to the probe.
To assemble the endocavitary biopsy bracket
Perform the following steps:
1. Ensure there is no damage, deformation, malfunction, loose or missing some parts on
the biopsy bracket before the assembly.
2. Wear sterile gloves.
3. Unfold the probe sheath, and apply an appropriate amount of coupling gel to the
inside of the sheath.
4. Hold the probe and unroll the sheath onto the probe.
5. Pull the probe sheath tightly over the face of the probe to remove wrinkles.
6. Assemble the metal top housing and biopsy bracket.
7. Slide the inner part into the top housing and push forward until it clicks.
Top Housing
Inner Part