20
User Manual
21
Intraoral Scanner i500
IEC 60601-1-2 Test Level Compliance Level
± 8 kV contact ± 2 kV, ± 4 kV, ± 8 kV,
± 15 kV air
± 8 kV contact ± 15 kV air
IEC 60601-1-2 Test Level Compliance Level
± 2 kV 100 kHz repetition frequency ± 2 kV 100 kHz repetition frequency
Mains power quality should be that of a typical commercial or hospital environment.
Electrical fast transient/burst IEC 61000-4-4
IEC 60601-1-2 Test Level Compliance Level
± 0.5 kV , ± 1 kV ± 1 kV
Mains power quality should be that of a typical commercial or hospital environment.
Surge Line-to-line IEC 61000-4-5
IEC 60601-1-2 Test Level Compliance Level
0 % UT; 0.5 cycle At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles Single phase: at 0°
0 % UT; 0.5 cycle At 0°, 45°, 90°, 135°,
180°, 225°, 270° and 315°
0 % UT; 1 cycle and 70 % UT; 25/30
cycles Single phase: at 0°
Mains power quality should be that of a typical commercial or hospital environment. If
the user of the EUT image intensier requires continued operation during power mains
interruptions, it is recommended that the EUT image intensifier be powered from an
uninterruptible power supply or a battery.
Voltage dips IEC 61000-4-11
IEC 60601-1-2 Test Level Compliance Level
± 0.5 kV, ± 1 kV, ± 2 kV ± 2 kV
Mains power quality should be that of a typical commercial or hospital environment.
Surge Line-to-ground IEC 61000-4-5
6 Electro – Magnetic Compatibility Information