22
User Manual
23
Intraoral Scanner i500
6 Electro – Magnetic Compatibility Information
IEC 60601-1-2 Test Level Compliance Level
3 V 0,15 MHz – 80 MHz 6 V in ISM
bands between 0,15 MHz and
80 MHz 80 % AM at 1 kHz
3 V 0,15 MHz – 80 MHz 6 V in ISM
bands between 0,15 MHz and 80 MHz 80 %
AM at 1 kHz
where P is the maximum output power rating of the transmitter in watts (W) according to the
transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site
survey, should be less than the compliance level in each frequency range.
Interference may occur in the vicinity of equipment marked with the following symbol:
IEC 60601-1-2 Test Level Compliance Level
3 V/m 80 MHz – 2,7 GHz 80 % AM at 1 kHz 3 V/m 80 MHz – 2,7 GHz 80 % AM at 1 kHz
Recommended separation distance
d = [ ]√P
7
E
1
d = [ ]√P
3,5
V
1
d = [ ]√P
3,5
E
1
80 MHz to 800 MHz
800 MHz to 2.5 GHz
Portable and mobile RF communications equipment, including cables, should be
used no closer to any part of the EUT than the recommended separation distance as
calculated using the equation below, according to the frequency of the transmitter.
Radiated RF IEC 61000-4-3