Weekly Maintenance
1. Clean CER reprocessor and basin
(use lint-free cloth)
2. Inspect CER reprocessor and hook-up components
for wear and tear
3. Lubricate O-rings with silicone oil
Monthly (as specied)
1. 1.0 micron water pre-lter (replace every 3 months
or if below 40 PSI)
2. 0.2 micron absolute lter (replace every 6 months
or if below 40 PSI) and perform waterline
disinfection
3. Active Vapor Management System (if applicable):
Charcoal lter (replace every 6 months)
MAINTENANCE
TROUBLESHOOTING
Troubleshooting
1. Review troubleshooting section in CER User
Manual
2. Service Contract Coverage and Loaner Repair
Program
Run the Reprocessing Cycle
1. Close CER lid and secure using lid locks
2. Cycle Selection: Automatic FULL or LCG/RINSE
3. Conrm Time Settings: Wash and LCG
4. Select START to begin cycle
5. Basin thermometer HLD temperature verication
Complete the Reprocessing Cycle
1. Verify STOP LED now illuminated
2. Alcohol Flushing Port w/air on (Manual, air/10, Start)
3. Record Endoscope Cycle completion in Log
4. Disconnect hook up and remove and store reprocessed endoscope(s) and hook up(s)
End of Day Shutdown
1. Water Supply Valve: End of Day “OFF”
2. Clean basin, lid and gaskets (Use lint-free cloth)
SYSTEM OPERATION
Medivators Website “Resource Center”
Go to: www.medivators.com, Select “Resource Center” and “User Library” for detailed user guides and hook up
matrices, report forms and logs, and product bulletins
Medivators Customer Service Support Toll Free: Phone: 1-800-444-4729 · FAX: 1-866-421-7696