Lists the components included in the 120V PM kit.
Lists the components included in the 230V PM kit.
Lists the tools necessary for flushing the SSG unit before maintenance.
Steps to prepare the SSG unit for maintenance, including idle state and lid placement.
Detailed steps for flushing the SSG unit's A side reservoirs using diagnostic commands.
Detailed steps for flushing the SSG unit's B side reservoirs using diagnostic commands.
Instructions for draining and rinsing the unit's basins after flushing procedures.
Steps to drain the uptake tube system to remove residual fluid.
Pre-maintenance preparation and parts list for 3/4-inch valve maintenance.
Maintenance procedures specifically for valves 9 and 11.
Maintenance procedures for valve 10, the sample port valve.
Steps for reinstalling the manifold after filter screen replacement.
Instructions for removing the detergent and alcohol mounting bracket on side B.
Mechanical steps involved in removing the side B manifold components.
Mechanical steps for removing the SSG dosing unit from the machine.
Guidance on using the O-ring removal tool for SSG reservoir O-rings.
Mechanical steps for replacing O-rings on the SSG reservoir pumps.
Steps to disconnect hydraulic lines for A side manifold removal.
Steps to disconnect electrical connectors for A side manifold removal.
Steps to disconnect hydraulic lines for B side manifold removal.
Steps to disconnect electrical connectors for B side manifold removal.
Steps for replacing valve seals, including removal and reassembly.
Steps for replacing check valves, noting orientation and O-ring removal.
Steps for replacing valve seals on alcohol and detergent valves.
Steps for replacing the disinfectant pump assembly on both A and B sides.
Steps for replacing peristaltic pump tubes on detergent and alcohol pumps.
Pre-reinstallation checks and mechanical steps for the A side module.
Pre-reinstallation checks and mechanical steps for the B side module.
Steps for reinstalling detergent and alcohol bottles after module reinstallation.
Steps for replacing check valves in the alcohol and detergent lines.
Instructions for replacing the basin sensor cover, noting critical hole orientation.
Steps to activate and prime alcohol and detergent pumps on both sides using diagnostics.
Checks for cycle parameters, sensor operation, and leak-free disinfection/rinse cycles.
Procedure to check and adjust rinse and disinfectant drain times in diagnostics.
Procedure to validate the operation of the internal leak tester using the service manual.
| Manufacturer | Medivators |
|---|---|
| Model | DSD EDGE |
| Type | Automated Endoscope Reprocessor |
| Number of Endoscopes per Cycle | 2 |
| Capacity | 2 endoscopes |
| Power Requirements | 120V, 60Hz, 15A |
| Drain Requirements | 1.5 inch drain line |
| Compliance | CE |