6
NNUAL PM INSTRUCTIONS
PREVENTATIVE MAINTENANCE KIT
Figure 2 Figure 3
Preparation
Close the external incoming water valve.
Bleed the 0.2-micron filter by pressing SETUP, 43, ENTER. This will allow the main water valve (#14) to open.
Place a small container under the bleeder hose to collect the water. Slowly open the bleeder valve until all flow
stops. Close the valve. Press STOP twice.
Unplug the DSD EDGE AC Power cord from wall outlet.
Air tank will be purged after the removal of the SSG Dosing Unit.
3/4–inch Valve Maintenance
Parts Required Part number QTY
Drain Rebuild Kit 78400-658 2
Overflow Rebuild Kit 78400-659 2
Strainer 43100-254 2
(one for each side)
O-ring MO01-0004 2
Tools Required
#2 Phillips screwdriver
9/64-inch Allen
3/16-inch Allen
Silicone grease