2
A
NNUAL PM INSTRUCTIONS
PREVENTATIVE MAINTENANCE KIT
230 Volt Kit 78401-177
PM Kit Part Content Part number QTY
Hydraulic Valve Seals 78401-261 1 (Kit of 10 seals)
Solenoid Valve Seal MV01-0035 4
Peristaltic Tube Set MT01-0500 4
Check Valves 78401-260 1 (Kit of 8 valves)
Pump Assembly 78400-685 2
Drain Rebuild Kit 78400-658 2
Overflow Rebuild Kit 78400-659 2
Sample Port Seal 41013-001 2
Strainer 43100-254 2
O-rings SSG Reservoir 43100-302 28
Check valve ball for SSG Res 47049-658 4
O-ring Removal Tool 47049-687 1
Instructions 50098-296 1
O-ring MO01-0004 2
Spring for SSG 47049-415 4
Sensor Cover MC13-0015 2
Procedure for flushing SSG unit prior to servicing
Before performing repairs on a SSG unit, the following procedure must be performed to flush Part A and B of the
RAPICIDE
®
PA solution from the unit. The flushing of the unit minimizes exposure to the disinfectant during repair or
preventative maintenance.
Tools Required
1/2 gallon or larger container for holding water. The vessel should be able to fit in the chemical drawer once the part
A and B bottles are removed. Two empty part A and B bottles can be used. If using empty bottles, rinse thoroughly
and remove the labels from the bottle to prevent confusion with actual disinfectant.
Table salt.