3
A
NNUAL PM INSTRUCTIONS
PREVENTATIVE MAINTENANCE KIT
PROCEDURE
CAUTION: PPE must be worn before proceeding. Refer to the MSDS for guidance.
Preparing the SSG Unit:
Both sides of the machine must be in an idle state.
The floating lids must be installed in both sides and the lids closed.
Fill the container 3/4 full of water. Add ten grams (2 teaspoons) of table salt for every 5 liters of water. If using the
empty bottles, add ten grams (2 teaspoons) of table salt to each filled bottle.
Remove the uptake tubes from both the part A and B bottles. Remove the bottles from the drawer.
Set the container of water in the drawer and set both uptake tubes in the container. See figure 1
Figure 1