Medivators DSD-201 - Reprocessor Leveling Procedure

Medivators DSD-201
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
^
|
e
>
GO
MEDIVA
SYSTEMS
Overriding
the
Disinfectant
Warning
The
Maximum
Disinfectant
Cycle
Count
is
preset
in
the
Diagnostics
menu.
After
the
disinfectant
reaches
10
cycles
before
the
maximum
number
of
cycles,
the
warning
indicator
illuminates
and
the
“Dis
Warning”
message
displays
on
the
LCD
screen.
The
warning
indicator
remains
illumi-
nated
until
either
the
disinfectant
is
changed
or
the
disinfectant
warning
override
(Setup
7)
is
ini-
tiated.
Use
the
procedure
below
to
override
the
Disinfectant
Warning.
Press
the
STATION
SELECT
button
to
choose
station
A
or
station
B.
Press
the
SETUP
button
on
the
reprocessor
control
panel.
Enter
7
on
the
keypad,
then
press
the
ENTER
button.
The
message
“Dis
Warn Ack”
is
displayed
on
the
screen.
SNP
Set
the
“Dis
Warn
Ack"
to
zero.
-
This
overrides
the
disinfectant
warning
for
another
10
cycles.
-
The
“Dis
Expired”
message
is
displayed
when
the
maximum
cycle
count
is
reached.
Warning!
Never
use
disinfectant
beyond
the
manufacturer’s
®
recommended
reuse
life,
even
if
the
MRC
level
is
><
acceptable.
®
Warning!
Never
use
disinfectant
that
is
below
acceptable
MRC
><
level,
even
if
the
reuse
date
is
unexpired.
126
MM03-0047
Revision
D