REPROCESSING
SYSTEMS
w
LEVELING
The
reprocessor
must
be
installed
on
a
level
surface
or
be
adjusted
to
level.
Adjust
the
lev-
eling
pads
after
unpacking
the
reprocessor.
134
1.
OOA
R
ON
Loosen
the
disinfectant
reservoirs.
On
older
models,
remove
the
2
bolts
securing
the
reservoir.
*
On
newer
models,
remove
the
2
pins
securing
the
reservoir.
Slide
the
reservoirs
forward
far
enough
to
access
the
rear
leveling
pads
(bolts).
Place
a
bubble
level
on
the
basin
surface
in
position
“A”.
Adjust
the
rear
leveling
pads
until
the
bubble
is
centered
in
the
level
glass.
Move
the
bubble
level
to
position
“B”.
Adjust
the
front
leveling
pads
(flathead
screws)
until the
bubble
is
centered
in
the
level
glass.
7.
Verify
the
reprocessor
is
level.
Slide
the
reservoirs
back
to
the
original
position
and
secure.