REPROCESSING
SYSTEMS
%
DISINFECTANT
FILTER
CARTRIDGE—-REPLACE
The
disinfectant
filter
must
be
changed
at
every
disinfectant
dump/load
procedure.
1.
Place
a
container
under
the
filter
to
catch
any
excess
liquid.
2.
Disconnect
the
quick-connect
fitting
from
the
reservoir
side
of the
filter.
Note:
Check
for
lint/debris
build
up
at
the
filter
connections.
Clean
the
filter
connections
before
installing
new
filter.
3.
Disconnect
the
quick-connect
fitting
from
the
pump
side
of
the
filter.
4.
Remove
the
filter
cartridge
and
replace
with
a
new
Medivators
cartridge.
*
Position
the
filter
with
the
arrows
in
the
direction
of
flow,
away
from
the
reservoir.
5.
Re-connect
the
quick-connect
fitting
to
the
pump
side
of
the
filter.
6.
Re-connect
the
quick-connect
fitting
to
the
reservoir
side
of
the
filter.
7.
Record
the
date
of
the
filter
change
in
the
log
book.
138
MM03-0047
Revision
D