Medivators DSD-201 - Air Filter Replacement Procedure

Medivators DSD-201
210 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
=
ğ
‘77
REPROCESSING
SYSTEMS
w
INTERNAL
WATER
FILTER-REMOVE
The
disinfector
must
be
in
idle
state
to
perform
this
procedure.
1.
Close
the
incoming
water
supply
valve
to
the
disinfector.
2.
Drain
excess
water
from
the
filter
housing.
*
Place
a
container
under
the
water
filter
inlet
tube.
*
Disconnect
the
water
filter
inlet
quick-connect
from
incoming
water
supply
line.
*
Connect
the
accessory
hose
to
the
water
filter
inlet
tube.
*
Open
the
filter
bleeder
valve
and
drain
the
water
from
the
filter
canister.
3.
Remove
the
water
filter
cartridge.
Warning!
Always
wear
gloves
when
handling
the
filter
cartridge.
a
Loosen
the
internal
filter
and
remove
the
water
filter
housing.
*
Rotate
the
filter
counterclockwise
to
unlock
the
filter
cartridge.
+
Remove
the
filter
and
discard
in
accordance
with
institution
guidelines.
140
MM03-0047
Revision
D