REPROCESSING
SYSTE
y
142
INTERNAL
WATER
FILTER-REPLACE
1.
Install
the
new
water
filter
cartridge.
Warning!
Always
wear
gloves
when
handling
the
filter
cartridge.
oS
+
Insert
the
filter
cartridge
into
the
housing
cap.
+
Turn
the
filter
cartridge
clockwise
until
the
tabs
locks
into
the
cap.
2.
Wipe
the
filter
housing
clean
with
a
lint-free
cloth.
3.
Install
the
filter
housing
onto
the
housing
cap.
ON
O
A
PR
*
Apply
an
NSF-approved
silicone
lubricant
to
the
housing
o-ring
to
aid
assembly
and
sealing.
Tighten
the
housing
into
the
cap
by
hand.
Do
not
overtighten.
Reconnect
the
water
inlet
line.
Close
the
bleeder
valve
on
the
filter
housing.
Turn
on
the
water
supply
and
check
for
leaks.
Bleed
any
air
from
the
housing
using
the
bleeder
valve.
Perform
the
Water
Line
Disinfect
procedure
as
described
in
the
Operator
Controls
chapter.
MM03-0047
Revision
D