PROBLEM ACTION
Unit does not
turn on
1. Check power source and connection
2. Ensure that the wall outlet is live
3. Check power cord is not damage
Motor runs, no
vacuum
1. Verify tubing connection security
2. Check for leaks or tubing kinks
3. Ensure that float shut-off is not activated
4. Check for bottle leaks and cracks
Low Vacuum 1. Use vacuum adjust knob to increase vacuum level
2. Check system for leaks
3. Adjust vacuum adjust knob and release
Checking system for leaks:
1. Detach all tubing from the machine and canister.
2. Turn on the HCS7000 and increase suction level to 21inHg
3. Occlude hose barb on the valve assembly (outlet on machine) and evaluate
suction level.
4. If level remains at 21inHg please continue to step 5. If level drops please contact
your Medline representative.
5. Add tubing with filter to machine and occlude furthest opening, again watching
suction level. If suction is lost please change filter and tubing and repeat.
6. Add canister and repeat suction test. If suction is lost please replace canister.
7. Continue building rest of your current setup until problem area is identified. Any
loss of suction please replace component that was last added to test circuit.
11. Disposables and parts for reorder:
Disposables
AGIAG405SET Hydrophobic Filter, Tubing and Elbow Assembly Pack 1 Set Pack
AGIG405SET3PK Hydrophobic Filter, Tubing and Elbow 3 Sets Pack
Assembly Pack
SUHBF4053 Replacement Hydrophobic Filter Only 3 Filter Pack
AGIBF405 Replacement Hydrophobic Filter Only 1 Filter Pack
HCS7850 Disposable 850 cc Suction Canister wit float Lid
HCS7851 Disposable 850 cc Suction Canister wit float Lid, with Suction
Tubing Kit
HCS7120 Disposable 1200 cc Suction Canister wit float Lid
HCS7121 Disposable 1200 cc Suction Canister wit float Lid, with Suction
Tubing Kit