30 MEDLINE | MDS4003
Table 4
Recommended separation distances between portable and mobile RF communications
equipment and the device.
The device is intended for use in an electromagnetic environment in which radiated RF
disturbances are controlled. The customer or the user of the device can help prevent
electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the device as recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output
power of transmitter
(W)
Separation distance according to frequency of transmitter
(m)
0.01 0.12 0.12 0.23
0.1 0.37 0.38 0.73
1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in metres (m) can be estimated using the equation applicable to
the frequency of the transmitter, where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency
range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
P P P
150 kHz to 80 MHz
d=3.5
80 MHz to 800 MHz
d=1.2
800 MHz to 2.7 GHz
d=3.5

Other manuals for Medline MDS4003

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Medline MDS4003 and is the answer not in the manual?

Medline MDS4003 Specifications

General IconGeneral
BrandMedline
ModelMDS4003
CategoryBlood Pressure Monitor
LanguageEnglish

Related product manuals