D-0126021-A 24
Safety
Regarding electrical safety, this device is designed to be used only by professionals in the hearing healthcare industry.
Regarding electrical safety, this device is to be used only by
professionals in the hearing healthcare industry.
It is Class II Medical Electrical (ME) equipment that is part of an ME
system.
Power is supplied by the USB cable connected to a computer. An in-
line optical isolator with medical power supply is recommended to
achieve optimum safety. A medical grade computer is recommended
and a medical grade isolation transformer will help minimize ground
current. It is recommended that powered speakers are connected to
an isolation transformer. The use environment should be between
10°C and 35°C, humidity within 30% to 90%. All components with
operator contact are made of bio-compatible materials.
The green “ready” light on the front of the device indicates that the
device and software are ready for use. Mild soapy water on a clean
cloth is the preferred cleaning solution.
The Instructions for Use (the Installation and Software Training
manuals) are supplied as an electronic copy on a USB flash drive. Paper
copies of the manuals may be also requested from the company, and
will be sent within one business day of the request.
All repairs should be sent to MedRx for evaluation and / or repair.
Symbols that may be used
or SN
Read the instruction manuals
for safe usage of the device
(operating instructions)
Indicates that the
device serial number
will follow
Type B applied part.
(Type B equipment)
Manufacturer
(MedRx)
Authorized
Representative in
Europe

Other manuals for MedRx AVANT HIT+