Compliance
Basic Safety and Essential EN 60601-1: 2006+A1:2013, EN 60601-1-2:2007
Performance
Type of Protection Internally Powered Equipment (on battery power)
Degree of Protection Non-Applied Part
Range Body mode: 93.2 °F - 109.2 °F (34.0 °C - 42.9 °C)
Surface mode: 32.0 °F – 212.0 °F (0.0 °C – 100.0 °C)
Accuracy Body mode:
93.2 °F - 94.8 °F ±0.5 °F
(34.0 °C - 34.9 °C ±0.3 °C)
95.0 °F – 107.6 °F ±0.4 °F (35.0 °C – 42.0 °C ±0.2 °C)
107.8 °F – 109.2 °F ±0.5 °F (42.1 °C – 42.9 °C ±0.3 °C)
Surface mode:
±3.6 °F (±2.0 °C)
Temperature Fahrenheit: 0.1
Resolution Celsius: 0.1
Three Color Backlight Green - Normal Temperature: 95.9°F – 99.2 °F
(35.5 °C – 37.3 °C)
Yellow - Slight Fever: 99.3 °F – 100.4 °F (37.4 °C – 38.0 °C)
Red - High Fever: 100.5 °F – 109.2 °F (38.1 °C – 42.9 °C)
Notes: 1.Surface Mode always displays a green backlight.
2. In Body Mode, a reading of 93.2 °F – 95.8 °F
(34.0 °C – 35.4 °C) displays a green backlight.
Auto Power O ≤ 18 s
Measuring Time ≤ 2 s
Measuring Distance 0.4 – 2.0 in (1 – 5 cm)
Memory 50 measurements
06
Power Supply Requirements
Batteries 1.5V AAA Alkaline Battery x 2 (IEC Type LR03)
Adaptable Range 2.6V – 3.6V
Environmental Conditions
Operating Conditions Temperature: 59.0 °F – 95.0 °F (15.0 °C – 35.0 °C)
Relative Humidity: 85%
Atmospheric Pressure: 70 – 106 KPa
Transport and Storage Temperature: -4.0 °F – 131.0 °F (-20.0 °C – 55.0 °C)
Relative Humidity: 93%
Atmospheric Pressure: 70 – 106 KPa
Conditions
Weight
(without batteries)
3.0 ounces (84 grams)
Size Length: 5.4” (138mm)
Width: 3.7” (95mm)
Height: 1.6” (40mm)
Physical Properties