OPERATING MANUAL
CO
2
incubators
INCO 108 INCO 108 med
INCO 153 INCO 153 med
INCO 246 INCO 246 med

Table of Contents

Question and Answer IconNeed help?

Do you have a question about the Memmert INCO 153 med and is the answer not in the manual?

Summary

Safety Regulations

Product Safety and Dangers

Details potential hazards like electric shock, suffocation, and burns from CO2 gas.

Accident Procedures and Emergency Shutdown

Steps for accidents, first aid, and emergency shutoff.

Design and Function

Intended Use and Restrictions

Specifies approved uses (cell cultures) and prohibits others (drying, sterilization).

Delivery, Transport, and Setting Up

Safety During Handling and Transport

Warns about injury risks during transport and installation.

Putting into Operation

Oxygen Calibration Procedure

Performing O2 calibration before initial operation.

Operation and Control

Basic Operation and Parameter Setting

Overview of the control panel and setting parameters.

Normal Mode Operation Details

Details permanent operation and setting parameters like temperature, humidity, and CO2.

Normal Mode Settings Example

Illustrates setting temperature and monitoring limits in Normal mode.

Humidity and Gas Setpoint Adjustment

Guides setting humidity, CO2, and O2 levels in Normal mode.

Programming Week Time Switch

Covers setting weekdays, day groups, and on/off times.

Week Time Switch Example

Provides a practical example of programming the week time switch.

Warning Messages and Malfunctions

General Warning Messages

Lists warnings like DOOR OPEN, TB active, HI-ALARM, LO-ALARM, ASF-ALARM, RH EMPTY.

Troubleshooting Common Errors

Provides a table of common errors, causes, and remedies.

Advanced Error Troubleshooting

Lists more specific errors like CONF message, fan issues, heating icon, and calibration errors.

Power Failure Handling

Advanced Functions

Temperature Monitoring System

Explains the Pt100 sensor and double overtemperature fuse.

Automatic Temperature Monitor (ASF)

How the ASF tracks setpoints within a tolerance band and triggers alarms.

Temperature Calibration Process

Customer-specific calibration using three temperature points.

Setting Temperature Calibration Values

Step-by-step guide to entering compensation values for temperature calibration.

Humidity Calibration Procedure

Customer calibration at 20% and 90% relative humidity.

CO2 Calibration Procedure

Customer calibration at 5%, 10%, and 15% CO2 levels.

Maintenance and Servicing

Memmert INCO 153 med Specifications

General IconGeneral
Temperature uniformity±0.3°C
Temperature stability±0.1°C
Power supply230 V, 50/60 Hz
Weight85 kg
CO2 control range0 to 20 %
Exterior dimensions (W x D x H)710 x 655 x 865 mm