W h en u sin g the tab le, ensure tha t a ll accessories are properly mounted to it and check the function of acces-
sory lockings and adjustments.
W hen using the s peciality headrest (100020126), ensure patient safe ty by relea sing the hea drest locking
before adjusting the back section. A warning label is affixed to the headrest.
The antistatic properties of the operating table require the use of Merivaara brand mattresses and antistatic
flooring.
Gas spring dismantling instructions are available from your sales representative. The release of nitrogen gas
is strictly
prohibited, without following the proper instructions.
If the operating table has been in the cold, it must be allowed to warm up at room temperature for at least 6
hours before recharging the battery or switching on, to allow any condensation formed to evaporate.
All pa cka g in g cardb oa rd sh ou ld be recycle d . Wood an d pla s tics are en erg y was te .
The op era ting tab le is he a vy (165 k g ).
We recommend recharging the battery regularly overnight after a days use. This ensures that the table is
always ready for use and it extends the battery life. Charging takes approx. 5-10 hours, depending on the
batte ry charg e sta tus . Th e ba tterie s al so have su fficie n t capa city to b e charg e d only on ce a we e k.
The eq u ip ote ntia l po int (DIN 42801) is locate d on the tab le ba se .
We recommend use of the equipotential point whenever operating and when using
patient monitoring equipment. Read all manufacturer instructions.
Dry the operating table carefully
by wiping down with a dry cloth immediately after cleaning or disinfecting.
If the operating table batteries should quit during a procedure, connect the table to the main power supply
and continue as normal. Batteries can be charged while adjusting the table.
The switch in the table base can also be used as an emergency shutoff, if necessary.
Becau se batteries are problem waste, they must taken to a problem waste disposal facility.
W hen moving the operating table, ensure that any accessories mounted on the side or the back section
kidney elevator crank do not cause any damage.
WARNINGS AND OBSERVATIONS
!
!
!
!
!
!
!
!
!
!
!
!
!
!
!