by inadequate staining and/or FISH (fluorescent in situ hybridization). Follow your evaluated
laboratory protocols and applicable SOPs.
Avoid Vibrations For prolonged experimental procedures using the microscope choose
an installation location with minimum vibration. Vibration stemming e. g. from forced ven-
tilation, floor impact or other shocks can lead to reduced image quality and experimental
errors. To reduce vibrations at the installation location, engage suitable damping measures.
Minimal space requirements The minimal area required for setting up your system is 100
cm x 60 cm [Metafer 150 cm x 60 cm] (width x depth). Observe the applicable national
regulations for VDU work stations and EN ISO 9241.
Power supply If the electricity is subject to local voltage fluctuations, we recommend the
usage of a UPS (uninterrupted power supply). Laser printers must NOT be connected to a
UPS due to their high initial current peak during warm up.
Computer network Several systems can be networked together using “Cat. 5” cables with
standard RJ-45 connections. The required data transfer rates are at least 100 Mbps. The
network installation is not included in the system price.
Changing system settings The system settings are preconfigured by MetaSystems or it’s
representatives. These settings should only be changed by MetaSystems personnel or by
experienced users. Pleased make a note of any changes made as some modifications may
cause severe software malfunctions Please consult the accompanying manuals and contact
MetaSystems (§).
Dust - Protect the microscope and camera from dirt and dust, which may impair with the
functionality of the camera and other system components. If you remove the camera from
the microscope, please ensure that the sensor is pointing downwards. The sensor is not
protected by a glass filter. Do not touch the sensor and do not try to clean the camera
yourself. Cameras should only be cleaned by the camera manufacturer or by MetaSystems
Cleaning Shut down all pieces of equipment belonging to the device and remove the
power cord of all system components prior to cleaning. Only use a soft damp cloth for
cleaning. Never employ liquid or spray cleaner containing inflammable substances.
Sensitive Fluorescence Filters If your microscope is equipped with fluorescence filters, you
should be aware that they are particularly sensitive to heat radiation. Prolonged exposure
to fluorescence light may cause irreversible damage leading to a complete loss of function-
ality. Ensure that proper heat protection filters are positioned in the light path between
lamp and fluorescent filters. Please follow the guidelines of the filter manufacturer with
regard to filter maintenance and cleaning (§).
Software updates Software updates are available free of charge for 1 year during the
initial free upgrade period after installation. You can update the software yourself following
the detailed installation instruction included. You may receive new software versions elec-
tronically or on CD-ROM.
Warming up of the Camera Cameras can get warm during operation. This may lead to a
poorer signal-to-noise ratio causing a reduction in image quality. Ensure a tight connection
between camera and microscope (camera adapter) to dissipate this heat energy through
the microscope.