EasyManua.ls Logo

METREL MI 3200 - Page 34

METREL MI 3200
46 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
MI 3200 TeraOhm 10 kV Measurements
34
Ustep ... Voltage step approx. 25 V (fixed value - not presetable)
Ustart .. Starting voltage
Tstep ... Test voltage duration per step
Tend .... Constant test voltage duration after reaching End value
t ........... Time
Ub ....... Breakdown voltage
Selecting this function displays the following states. The Fig. 28 shows the initial screen
and a screen with results after the completion of a measurement.
Initial display
Display with results
Fig. 28. Withstanding Voltage function display states
Legend of displayed symbols:
WITHSTANDING VOLTAGE DC
Name of selected function
2000V
Start test voltage
7000V
Stop test voltage
7221V
Actual test voltage – measured value
I=0.002mA
Actual test current – measured value
tm:01min 00s
Timer information
Measuring procedure:
- Connect the test leads to the instrument and to the measured object.
- Press the START/STOP key to start the measurement.
- Wait until the set timers run out or until breakdown occurs, (the result will be
displayed).
- . Wait for the object under test is discharge.
- The result can be saved by pressing the MEM key twice, see the chapter 6.1 Store,
Recall and Clear Operation.
Note:
- Breakdown is detected when the measured current reaches or exceeds the set
current level Itrig.
Notes:
- The timer shows the time needed to complete each step during the measurement
and it shows the total measurement period after the completion of the measurement.
- A high-voltage warning symbol appears on the display during the measurement to
warn the operator of a potentially dangerous test voltage.